The UMS2 fully automated scanning system has been specifically designed to automate the repetitive tasks associated with the acquisition, display, storage and computation of data associated with the measurement and mapping of acoustic fields.
Unlike our competitors we do not make a standard product. ALL of our test tank systems are customised to suit the users particular measurement situation. The information and pictures will give an indication of what is possible but please contact us with your specific requirements. The UMS2 scanning system has passed the tests for Radiated and Conducted Emmissions and Immunity and is therefore ideal for use in a research enviroment. | Ultrasound measurements rely on the ability to measure accurately the acoustic field.The UMS2 system has three motorised axes (x,y,z) each capable of a six micron resolution and is supplied complete with electronic controle module, PC with preloaded UMS software and oscilloscope. Options include: "3 rotational axes positioning device" facilitates accurate alignment of the transducer's acoustic axis with the hydrophone; water conditioning system; hydrophone and transducer mounts
|
Typical Applications:
- Beam profiling
- Location of focal area
- Pressure mapping
- Measurement of acoustic parameters to satisfy the latest regulatory requirements needed for IEC62127-1, NEMA UD3 or FDA 510(k) Track 111 submissions
The PA Test Tank System
- Automatic precise positioning of the PA hydrophone mount to ensure accuracy of the acquired data
- Customisation for individual clients but the supplied software controls the oscilloscope and stepper motor driver and allows the user to conduct linear and planar scans
- Fitted with encoders to ensure that the hydrophone mount returns accurately to its pre-defined position
The Supplied Software
- Integrates the positioning system, oscilloscope and PC for data storage
- Automates planar scanning routines
- The included post processing software enables the calculation of standard acoustic reporting parameters from the acquired data
- LabView source code for all functions is supplied with all sytems (requires LabView 8)
Specification
- Maximum tank dimensions: length - 1000mm, width - 450mm, height - 450mm.
- Scanning area: length - 600mm, width - 250mm, height - 250mm, increment - 6.25um
- Alternative geometry can be supplied
- Frequency range 0.5MHz - 20MHz
- LabView 8 www.ni.com
- Software integrates with Lecroy Oscilloscope (6000 Series)
Optional Extras
- Water Treatment system
- Perspex test tank to suit the supplied scanning system
- PA membrane or needle hydrophones
- Hydrophone mount for hydrophones from other manufacturers
- Alternative transducer mounts
- Function generators etc.
For more information, contact: technical@acoustics.co.uk