Products

Home / Products / Pressure / Intensity Measurements / Needle Hydrophones / 0.2mm probe

0.2mm probe

HP 0.2mm Interchangeable Probe - Order Code HPM02/1

All dimensions are in mm. All probes are hand made and as such there is always some variation between individual devices. Please contact us if the actual dimensions are critical to your work.

Description
The 0.2mm needle hydrophone is our best selling product. With good signal-to-noise ratios Precision Acoustics would recommend the use of this device in most measurement situations to minimise the effect of spatial averaging.

Specification

Output Impedance:
13pF+/- 2pF
Typical Probe Sensitivity:
55nV/Pa (equivalent to -265.2dB re 1V/uPa) at 3MHz when measured at the output of a correctly terminated HP1 Submersible Preamplifier
Sensitivity Tolerance:
Sensitivity variation between individual probes is approximately +/-3dB
Typical Frequency Response:
(as defined by IEC 60866)
Class A (+/-2dB): 5 to 25MHz
Class B (+/-4dB): 1 to 35MHz
Sensor Material:
9 micron thick gold electroded Polyvinylidene fluoride (PVdF) film
Sensor Dimension:
0.2mm diameter disc

This specification is subject to change without notice.

For further details contact one of our Applications Engineers on 01305 264669.